We are remarkable name in this domain, engrossed in offering top-notch grade Uv Visible Spectrophotometer. This spectrophotometer is manufactured using quality tested electronic components by adept crew of professionals with the help of contemporary techniques. The provided spectrophotometer is highly used in analytical chemistry for quantitative determination of different analytes such as transition metal ions and biological macromolecules. Furthermore, we put forward this Uv Visible Spectrophotometer at reasonable prices.
Features:
UV-1800
SHIMADZU UV-VIS SPECTROPHOTOMETER MANUFACTURED
BY SHIMADZU CORPORATION JAPAN.
A compact, double beam UV-VIS spectrophotometer wrapped in a sleek form. The UV-1800 uses the Czerny-Turner mounting for its monochromator, and boasts the highest resolution in its class, a bright optical system and a compact design. Available as either a Stand-alone instrument or a PC-controlled instrument. The UV-1800 is USB-memory ready, which enables users to save measurement date to highly versatile USB memory, and perform data analysis and printing using a PC. Supplied with 1 Pair of 10 mm rect. quartz Cell.
Features:
Specifications:
Wavelength range : 190 to 1100nm
Spectral bandwidth : 1 nm (190 to 1,100nm)
Wavelength display : 0.1 nm increments (1nm increments when setting scanning Range)
Wavelength accuracy : +/- 0.1 nm at D2 peak 656.1nm +/- 0.3 nm for entire range
Wavelength repeatability : +/- 0.1 nm
Wavelength slew rate : About 6000 nm/min
Wavelength scanning speed : 3000 to 2 nm/min Lamp interchangeable
Wavelength : Automatic interchangeable linked to wavelength can be set Freely in the range of 295 to 364 nm(0.1 increment)
Stray light : Less than 0.02% at 220nm (Nal) Less than 0.02% at 340 nm (NaNO2) Less than 1.0% at 198 nm (KCl)
Photometric system : Double beam optics
Photometric range : Absorbance: -4 to 4 Abs,
Transmittance : 0% to 400%
Photometric accuracy : +/- 0.002 Abs at 0.5Abs
+/- 0.004 Abs at 1.0 Abs
+/- 0.006 Abs at 2.0 Abs (measured using NIST 938D/NIST1930
or equivalent)
Photometric repeatability : Less than +/-0.001 Abs at 0.5 Abs
Less than +/- 0.001 Abs at 1 Abs
Less than +/- 0.003 Abs at 2 Abs
Baseline Stability : Less than 0.0003Abs/Hr (700nm 1hour after light source turned
ON)
Baseline flatness : Less than +/-0.0006 Abs (1100 - 190nm 1 hr after light source
Turned ON)
Noise level : Less than 0.00005 Abs (700 nm)
Light source : 20-W halogen lamp and D2 lamp built-in source auto position
Adjustment
Monochromator : Blazed holographic grating Czerny-Turner mounting
Detector : Silicone Photodiode
Sample compartment : Internal dia: 110(W) x250(D) x115(H) mm
Distance between light beams : 100mm
Environmental requirements : Temperature : 15°C to 35°C,
Humidity : 30% to 80%
Dimensions : 450 (W) x490(D) x270(H) mm
Weight : 15 kg.
Output device : USB memory (optional) Data files saved in text format or UVPC
Format, UVPC format files can be read directly by UVProbe
PC compatibility : Provided with UVProbe software, External control possible via USB
"We are mainly dealing with bulk quantity orders in local areas."